Abstract

This part of YY/T 0988 includes stereological testing methods for characterizing the coating thickness, porosity, and average intercept of various porous coatings adhered to non-porous substrates.
Section 9.4 outlines the measurement methods for porosity and intercept on the "tissue interface gradient" at different levels of porous coating thickness.
The tangential sampling method in Section 8.2 is not applicable to the tissue interface gradient method in Section 9.4.
This part adopts the International System of Units (SI).
This part does not attempt to address all safety issues involved, even those related to its use. It is the responsibility of the users of this standard to establish appropriate safety and health regulations and to clarify the applicability of management restrictions before use.