T/CIS 13003-2024 The determination of ammonia and hydrogen chloride in pollutant gas from fixed sources using portable tunable diode laser absorption spectroscopy
T/CIS 13003-2024 The determination of ammonia and hydrogen chloride in pollutant gas from fixed sources using portable tunable diode laser absorption spectroscopy
Basic Information
Scope
Scope: This document specifies the portable tunable semiconductor laser absorption spectroscopy method for measuring ammonia (NH3) and hydrogen chloride (HCl) in the exhaust gas of fixed pollution sources. This document is applicable to the determination of NH3 and HCl in the organized emissions of exhaust gas from fixed pollution sources. Note: - The detection limit of the portable tunable semiconductor laser absorption spectroscopy method: NH3 is not more than 0.3 mg/m3, and HCl is not more than 0.2 mg/m3; - The lower limit of determination: NH3 is not more than 1.2 mg/m3, and HCl is not more than 0.8 mg/m3; - Main technical content: - Table of Contents - Preface - Scope - Normative References - Terms and Definitions - Method Principle - Standard Gases and Preparation - Instruments and Equipment - Composition - Functional Requirements - Performance Requirements - Samples - Analysis Steps - Test Preparation - Instrument Verification - Sample Determination - Result Calculation and Presentation - Result Calculation - Result Presentation - Precision and Accuracy - Precision - Accuracy - Quality Assurance and Quality Control - Attention Points - Appendix A (Normative Appendix): Precision and Accuracy - Appendix B (Informative Appendix): Instrument Performance Audit Results Before and After Determination