T/ACEF 200-2025 Active Group standards

T/ACEF 200-2025 Quality Control Test Specification for Multi-Leaf Collimator of Mechanical Arm Radiotherapy Device

T/ACEF 200-2025 Quality control inspection specifications for multi-leaf collimators used in robotic radiation therapy devices

Publish Date: 2025-05-15 Implement Date: 2025-05-20 For services related to genuine standard inquiry, procurement, translation, and other related services in China, please Contact Us
Price: $ 20.00
Out of stock

Basic Information

Standard Code: T/ACEF 200-2025
Standard Type: Group standards
Standard Status: Active
is_force_gb: no
CCS Name: Medicine, health, and labor protection
ICS Name: Radiation protection
Publish Date: 2025-05-15
Implement Date: 2025-05-20

Scope

Scope: This document specifies the quality control inspection items, inspection methods, and requirements for multi-leaf collimators of robotic arm radiotherapy devices. This document is applicable to the quality control inspection of multi-leaf collimators of robotic arm radiotherapy devices. Main technical content: Quality control inspection items and methods for multi-leaf collimators of robotic arm radiotherapy devices include: 1. Deviation of irradiation field size; 2. Deviation of automatic quality assurance (AQA) for multi-leaf collimators; 3. Deviation of end-to-end (E2E) for multi-leaf collimators; 4. Test of adjacent rectangular irradiation fields; 5. Test of blade positioning accuracy; 6. Test of leakage radiation rate through the collimator.