GB/T 25917.1-2019 Active National standards

GB/T 25917.1-2019 Uniaxial fatigue testing systems—Part 1:Calibration of dynamic force

GB/T 25917.1-2019 Uniaxial fatigue testing systems—Part 1:Calibration of dynamic force

Publish Date: 2019-10-18 Implement Date: 2020-05-01 For services related to genuine standard inquiry, procurement, translation, and other related services in China, please Contact Us
Price: $ 66
Download

Basic Information

Standard Code: GB/T 25917.1-2019
Standard Type: National standards
Standard Status: Active
is_force_gb: no
CCS Name: Metallic material testing machine
ICS Name: Mechanical testing
Publish Date: 2019-10-18
Implement Date: 2020-05-01
Pages: 20 pages

Scope

This part of GB/T 25917 describes two methods for determining the relationship between the dynamic force range (ΔFt) of a specimen during a uniaxial, sinusoidal, constant amplitude test and the force value display range (ΔFi) of the test system (see Introduction).
These methods enable the dynamic test system to operate without encountering the system's resonance frequency, and are suitable for test systems where the dynamic force measurement error is unknown or where the applied force range is expected to exceed 1%.
The dynamic force measurement error is determined by comparing the peak value of the force displayed by the dynamic force test system with the measurement value of the dynamic force calibration device (DCD) equipped with a resistance strain gauge. A static calibration of the DCD's test system display value must be performed in advance (see 5.2.1).
The dynamic calibration of Method A (equivalent specimen method) is only applicable to the effective frequency range confirmed by the specimen. A frequency-based correction factor is used to correct the dynamic force measurement error up to 10% of the dynamic force range. Using this correction factor, the actual dynamic force measurement error in the test can be reduced to less than 1% of the dynamic force range.
The dynamic calibration of Method B (stiffness envelope method) is suitable for specimens with an effective frequency range between two sets of DCDs. Because Method B does not allow a dynamic force measurement error exceeding 1% of the dynamic force range, no correction factor is applied in Method B.
Note: Appendix A provides guidance on recalibrating the test machine using the methods described in this part.

Development Information

Word Count: 30 Thousand words Pages: 20 pages

Same series standard

Replace the following standards

Referenced Standards

ISO 4965-2 ISO 7500-1

Adopt standards

ISO 4965-1:2012

Related Standards