GB/T 21649.1-2008 Replaced National standards

GB/T 21649.1-2008 Particle size analysis—Image analysis methods—Part 1:Static image analysis method

GB/T 21649.1-2008 Particle size analysis—Image analysis methods—Part 1:Static image analysis method

Publish Date: 2008-04-16 Implement Date: 2008-10-01 For services related to genuine standard inquiry, procurement, translation, and other related services in China, please Contact Us

Basic Information

Standard Code: GB/T 21649.1-2008
Standard Type: National standards
Standard Status: Replaced
is_force_gb: no
CCS Name: Screening, sieve plates, and sieve nets
ICS Name: Granularity analysis and sieving
Publish Date: 2008-04-16
Implement Date: 2008-10-01
Pages: 36 pages

Scope

This part of GB/T 21649.1 specifies the method for determining the particle size distribution using static image analysis. This part is applicable to the determination of the particle size distribution of particles that can be obtained from a microscope. The particle size range measured by this part is a narrow distribution of less than 10:1. To obtain a reproducible volume average diameter, a distribution with a standard deviation of 1.6 that conforms to a normal distribution requires measuring more than 6,000 particles. If you want to obtain reliable percentage values, such as D90 or other percentage values, you must measure at least 61,000 particles. For details, see Appendix A.

Development Information

Word Count: 60 Thousand words Pages: 36 pages

Superseded by the following standards

Referenced Standards

ISO 9276-1 ISO 9276-2

Adopt standards

ISO 13322-1:2004

Related Standards

Contact Us