YY/T 0722-2016 Active Industry standards-Medicine

YY/T 0722-2016 Medical electrical equipment—Dosimetric instrument used for non-invasive measurement of X-ray tube voltage in diagnostic radiology

YY/T 0722-2016 Medical electrical equipment—Dosimetric instrument used for non-invasive measurement of X-ray tube voltage in diagnostic radiology

Publish Date: 2016-07-29 Implement Date: 2017-06-01 For services related to genuine standard inquiry, procurement, translation, and other related services in China, please Contact Us

Basic Information

Standard Code: YY/T 0722-2016
Standard Type: Industry standards
Standard Status: Active
is_force_gb: no
CCS Name: Medical radiation equipment
ICS Name: X-ray imaging equipment
Publish Date: 2016-07-29
Implement Date: 2017-06-01
Pages: 30 pages

Scope

This standard specifies the performance requirements and related conformity tests for non-accessible measuring instruments used for X-ray tube voltages below 150 kV. The standard also describes calibration methods and provides guidelines for estimating uncertainties under different measurement conditions during calibration.
Measurement applications in diagnostic radiology include breast imaging, computed tomography (CT), dental radiology, and X-ray examinations. This standard does not address the electrical safety requirements for the safe use of such instruments; these requirements are specified in GB 4793.1-2007.

Development Information

Word Count: 54 Thousand words Pages: 30 pages

Replace the following standards

Referenced Standards

GB 4793.1-2007 Safety requirements for electrical equipment for measurement, control, and laboratory use—Part 1:General requirements IEC 60417(所有部分) IEC 60788 IEC 61187:1993 ISO Guide 99:1993 ISO 7000:1989 GB/T 17626.2-1998 Electromagnetic compatibility—Testing and measurement techniques—Electrostatic discharge immunity test GB/T 17626.2-2006 Electromagnetic compatibility—Testing and measurement techniques—Electrostatic discharge immunity test GB/T 17626.2-2018 Electromagnetic compatibility—Testing and measurement techniques—Electrostatic discharge immunity test GB/T 17626.3-1998 Electromagnetic compatibility—Testing and measurement techniques-Radiated, radio-frequency, electromagnetic field immunity test GB/T 17626.3-2006 Electromagnetic compatibility—Testing and measurement techniques—Radiated, radio-frequency, electromagnetic field immunity test GB/T 17626.3-2016 Electromagnetic compatibility—Testing and measurement techniques—Radiated,radio-frequency,electromagnetic field immunity test GB/T 17626.3-2023 Electromagnetic compatibility—Testing and measurement techniques—Part 3:Radiated,radio-frequency,electromagnetic field immunity test GB/T 17626.4-1998 Electromagnetic compatibility—Testing and measurement techniques—Electrical fast transient/burst immunity test GB/T 17626.4-2008 Electromagnetic compatibility—Testing and measurement techniques—Electrical fast transient/burst immunity test GB/T 17626.4-2018 Electromagnetic compatibility—Testing and measurement techniques—Electrical fast transient/burstimmunity test GB/T 17626.5-1999 Electromagnetic compatibility—Testing and measurement techniques—Surge immunity test GB/T 17626.5-2008 Electromagnetic compatibility—Testing and measurement techniques—Surge immunity test GB/T 17626.5-2019 Electromagnetic compatibility—Testing and measurement techniques—Surge immunity test GB/T 17626.6-1998 Electromagnetic compatibility—Testing and measurement techniques—Immunity to conducted disturbances, induced by radio-frequency fields GB/T 17626.6-2008 Electromagnetic compatibility—Testing and measurement techniques—Immunity to conducted disturbances,induced by radio-frequency fields GB/T 17626.6-2017 Electromagnetic compatibility—Testing and measurement techniques—Immunity to conducted disturbances,induced by radio-frequency fields GB/T 17626.11-1999 Electromagnetic compatibility—Testing and measurement techniques—Voltage dips, short interruptions and voltage variations immunity tests GB/T 17626.11-2008 Electromagnetic compatibility—Testing and measurement techniques—Voltage dips,short interruptions and voltage variations immunity tests GB/T 17626.11-2023 Electromagnetic compatibility—Testing and measurement techniques—Part 11:Voltage dips, short interruptions and voltage variations immunity tests for equipment with input current up to 16 A per phase