YY/T 0482-2010 Magnetic resonance equipment for medical imaging—Part 1:Determination of essential image quality parameters
YY/T 0482-2010 Magnetic resonance equipment for medical imaging—Part 1:Determination of essential image quality parameters
Basic Information
Scope
This standard specifies the measurement procedures for determining the image quality parameters of many major medical magnetic resonance (MR) equipment. The measurement procedures described in this standard are applicable to:
——Quality evaluation during acceptance testing;
——Quality assurance during stability testing.
Additionally, the measurement procedures specified in this standard are also helpful for type testing, although this is not the purpose of this standard.
This standard is not applicable to:
——Determining the performance level during acceptance testing and stability testing;
——Image quality evaluation of high-field MR equipment greater than 4T, unless otherwise specified;
——Image quality affected by MR compatibility issues;
——Special diagnostic procedures, such as blood flow imaging, perfusion, diffusion, radiotherapy, and image-guided therapy applications;
——Type testing.
The scope of this standard is also limited to measuring the image quality characteristics of test molds, rather than patient images.