YY/T 0946-2014 Cardiac defibrillators—Connector assembly DF-1 for implantable defibrillators—Dimensions and test requirements
YY/T 0946-2014 Cardiac defibrillators—Connector assembly DF-1 for implantable defibrillators—Dimensions and test requirements
Basic Information
Scope
This standard applies to the DF1 unipolar connector assembly used when connecting implantable cardiac defibrillation electrode leads to implantable cardiac defibrillation pulse generators with a maximum output of no more than 1 kV/50 A. This standard specifies the basic dimensions, performance requirements, and corresponding test methods for the compatibility between components.
This standard does not specify other characteristics of the connector, such as fastening methods and materials.
This standard applies to the appearance and compatibility of the connector assembly.
This standard does not apply to the functional compatibility and system performance reliability of different implantable defibrillation electrode leads when used with implantable cardiac defibrillation pulse generators.